West Roxbury, Massachusetts, United States
Boppli® BP values compared to intermittent NIBP measurements during transport and after transport for one subject. Comparison of transport period and post-transit period demonstrated closer accuracy of the device in transit for SBP, but less accurate results for DBP and MAP, although still within acceptable limits for FDA standards.
Bland Altman analysis for subject TUF-029 from Figure 1 demonstrating that mean bias and standard deviation for transport (black dots) and post-transit (gray dots) are within FDA targets (green lines) for accuracy at 5 mmHg for SBP, DBP and MAP. Red lines represent 2-sigma values.